BRIEF COMMUNICATION  
Year : 2022  |  Volume : 15  |  Issue : 2  |  Page : 154-159
Choosing an appropriate size valve for transcatheter pulmonary valve implantation in a native right ventricle outflow tract


1 Department of Pediatric Cardiology, Apollo Children's Hospital, Chennai, Tamil Nadu, India
2 Department of Cardiothoracic Surgery, Apollo Children's Hospital, Chennai, Tamil Nadu, India

Click here for correspondence address and email

Date of Submission09-Apr-2021
Date of Decision18-Jun-2021
Date of Acceptance23-Aug-2021
Date of Web Publication19-Aug-2022
 

   Abstract 


Introduction: Transcatheter pulmonary valve implantation has been an effective treatment for dysfuntional right ventricular tract outflow tract (RVOT). Defining a landing zone before the intervention is crucial in patients with native RVOT. Improper sizing and undefined landing zone will lead to embolization. Methods: It is a retrospective observational study from August 2020 to December 2020 in native RVOT. Three patients who had significant Right ventricle dilatation were analyzed. The multi-slice computed tomography (MSCT) with magnetic resonance imaging and angiography data of all patients before the procedure were analyzed. All patients underwent an angiogram in the same sitting, before the procedure to assess the landing zone, valve diameter as well as the risk for coronary compression. We chose a valve based on valve area 23%–25% more than the area at the waist during balloon sizing. Results: All three patients underwent successful valve implantation. Valve sizes used were 27.5 mm in one and 32 mm in the other two. The mean RVOT gradient postprocedure was 11.5 mm Hg and pre procedure was 43 mmHg. There were no complications during the procedure or at a mean follow-up of 3.6 months. Conclusion: The balloon sizing gives the true narrowest diameter in comparison with MSCT, and increasing this area by 23%–25% will give the appropriate valve size for successful implantation.

Keywords: Congential heart disease, adults, percutaneous intervention, pulmonary valve disease, structural heart disease intervention, transcatheter valve implantation

How to cite this article:
Sivaprakasam MC, Reddy J R, Ganesan R, Sridhar A, Solomon N, Moosa MJ, Lakhani ZS, Gunasekaran S. Choosing an appropriate size valve for transcatheter pulmonary valve implantation in a native right ventricle outflow tract. Ann Pediatr Card 2022;15:154-9

How to cite this URL:
Sivaprakasam MC, Reddy J R, Ganesan R, Sridhar A, Solomon N, Moosa MJ, Lakhani ZS, Gunasekaran S. Choosing an appropriate size valve for transcatheter pulmonary valve implantation in a native right ventricle outflow tract. Ann Pediatr Card [serial online] 2022 [cited 2022 Sep 25];15:154-9. Available from: https://www.annalspc.com/text.asp?2022/15/2/154/354059





   Introduction Top


Transcatheter pulmonary valve (PV) implantation (TPVI) has been proven to be an effective treatment option for dysfunctional native right ventricular tract outflow tract (RVOT).[1],[2] Since the first report in 2000,[3] several case series have been published using the Melody® (Medtronic Minneapolis, MN, USA) and the Sapien® S3 (Edwards Lifesciences, Irvine, CA).[4],[5]

In most of the case series, prestenting was done in patients who require valve implantation to create a reliable and solid landing zone.[6],[7] Recently, publications showed good results with TPVI in native RVOT.[8],[9],[10] The Sapien® S3 valve was used and the largest available valve was 29 mm. A 29 mm valve becomes unsuitable for hugely dilated RVOT (most common). We reported the first TPVI using Myval® in a pulmonic position with prestenting.[11] Recently, Myval® (Meril Life sciences, Surat, India) transcatheter heart valves (THV) introduced 30.5 and 32 mm valve sizes. As there is a large spectrum of valve sizes available, we preferred them in dilated RVOT. We report our experience with three successful TPVI in the native pulmonary artery.


   Methods Top


Study design and patient population

It is a retrospective study. Three patients who had primary ICR for TOF were included from August 2020 to December 2020 [Table 1]. The symptoms of patients, magnetic resonance imaging (MRI) right ventricle (RV) volume, and PR were in favor of TPVI.[12] Informed written consent was obtained from all the patients.
Table 1: Baseline clinical characteristics of the study population

Click here to view


Preprocedural evaluation

All the patients underwent clinical examination, routine blood testing, electrocardiography, chest radiography, and echocardiography. Multi-slice computed tomography (MSCT) with MRI was performed to assess RVOT diameter, anatomy, and coronary course. MRI was also used to quantify ventricular volumes and function and PR fractions. All patients underwent Angiogram with balloon sizing in the same sitting, before the procedure to assess the landing zone, valve diameter as well as the risk for coronary compression [Figure 1]. Valve size used was based on area measurements on MSCT and angiogram [Figure 2], [Figure 3], [Figure 4], [Figure 5].
Figure 1: (a and b) Pigtail angiogram in right anterior oblique 30° cranial 30° and in lateral views showing dilated right ventricle and main pulmonary artery. (c and d) showing balloon sizing of main pulmonary artery segment along with simultaneous right ventricle and left anterior descending coronary angiogram using pigtail and an end hole catheter respectively. (e) The positioning of the valve across the defined landing zone simultaneous rapid pacing of the left ventricle with a pacing catheter already placed. (f) Pigtail MPA angiogram postvalve deployment with no paravalvular leak or valvular regurgitation

Click here to view
Figure 2: (a-c) Multislice computed tomography images showing delineation of right ventricular, right ventricular outflow, pulmonary valve, and the pulmonary artery to identify landing zone

Click here to view
Figure 3: (a-c) Multi-slice computed tomography three-dimension correlation of identifying narrowest zone for implantation and narrowest area measurements

Click here to view
Figure 4: Multi-slice computed tomography images (a) Calculation of the narrowest landing zone dimensions, perimeter, and area. (b) Calculation of the ostioproximal left pulmonary artery dimensions, perimeter, and area. (c) Calculation of the ostioproximal right pulmonary artery dimensions, perimeter and area

Click here to view
Figure 5: (a-c) Multi-slice computed tomography image relation of proximal left anterior descending coronary artery to the landing zone

Click here to view


Procedure

The procedure was done under general anesthesia. A right internal jugular vein and right radial artery access were obtained for monitoring. The right femoral vein was the primary access point for valve implantation. Intravenous heparin was administered to achieve an activated clotting time of >250 s. All patients were administered antibiotic prophylaxis during the procedure and for the 48 h following the procedure.

RV angiogram was performed in the right anterior oblique 30° (cranial 30°) and in lateral views [Figure 1]. The left femoral arterial access was obtained and a 6F sheath was positioned for selective coronary angiography. We accessed the coronary near to the landing zone and kept a guide catheter with a coronary wire. The left femoral venous access was obtained for RV angiogram. The right femoral artery access was obtained for pacing catheter in the LV. The right femoral sheath is upsized to the 14F Python™ sheath (Meril Life Sciences Pvt. Ltd., India) which is suitable for all the sizes of Myval® THV. Then, a Lunderquist wire was positioned deep in the left pulmonary artery. We used a 30 mm × 50 mm Z-MED™ balloon (NuMed™ Canada Inc) to size the RVOT. On rapid pacing at 250/min, we inflated the balloon and looked for a waist to appear. Then, we did simultaneous RV and coronary angiogram [Figure 1]c and [Figure 1]d. We proceeded with the valve implantation only when, we saw a minimal waist on balloon inflation, there was complete occlusion of RVOT without any forward flow and no coronary compression.

The waist diameter was noted and then we proceeded with TPVI. We chose a valve based on valve area 23%–25% more than the area at the waist during balloon sizing. The valve is manually crimped on the Navigator™ THV delivery system. The Navigator™ THV delivery system was advanced through Python™ sheath and the valve was positioned across the defined landing zone. An angiogram was done to confirm the position [Figure 1]e. Then, heart was rapidly paced at 250 beats/min. After the initial angiogram on pacing, the balloon was inflated slowly. Midway through the inflation, we performed another angiogram to reconfirm the position, followed by full inflation. A selective coronary angiogram was done to make sure there is no coronary obstruction. After deployment pulmonary artery (PA) angiogram was performed to assess the functioning of the valve [Figure 1]f. A catheter gradient across the valve was documented. Postprocedure echocardiography confirmed the valve position, RVOT gradient, RV function, and pericardial effusion.


   Results Top


All the three patients underwent successful TPVI with Myval+ without prestenting. The mean procedural time from placement to the removal of the sheaths was 95 min (range 90–100 min). The mean fluoroscopic time and dose area product were 26 min (range 24–28 min) and 75.2 Gy.cm2 (range 49.3–109.5 Gy.cm2), respectively. Valve sizes used were 27.5 mm in patient #1, 32 mm in patient #2 and #3. Diluted contrast saline was taken to inflate the balloon which contains a nominal volume of 28 ml for patient #1 who required a 27.5 mm valve. A 40 ml diluted contrast was recommended for the 32 mm valve which we implanted for patients #2 and #3. However, we have taken additional 2 ml and 5 ml to nominal volume for them, respectively. Dual antiplatelet agents were started and continued for 6 months. The mean follow-up duration is 3.6 months (range 2 to 5 months). Symptom status improved to New York Heart Association (NYHA) II (compared to pre-procedure NYHA III). Mean QRS duration postprocedure was reduced to 120 ms (compared to preprocedure 153 msec). The mean RVOT gradient postprocedure was 11.5 mmHg, compared to preprocedure was 43 mmHg [Table 2]. No stent fractures were seen in Chest X-ray which was taken 2–3 months postprocedure. None had valvular regurgitation or para-valvular leak postprocedure or on follow-up.
Table 2: Pre- and post-procedural characteristics of study population

Click here to view



   Discussion Top


In this case series, we have demonstrated that Myval® can be implanted successfully across the native PA. Myval® offers enough grip to be implanted without a pre-stent [Figure 6]. Previously, Tanase et al.[8] have demonstrated that Sapien® S3 can be safely implanted in native PA. However, the limitation of the Sapien® S3 is the largest valve for implantation available is 29 mm.[13] However, the majority of native PA is dilated and would require a large diameter than 29 mm.[14],[15] The largest available Myval® comes with a 32 mm valve with an outer diameter of 33 mm which gives a large spectrum for PV implantation.[16] Additional diluted contrast saline was required to add 2 ml and 5 ml in patient # 2 and # 3, respectively, to the nominal volume to increase the diameter.
Figure 6: Myval® (Meril Life Sciences Pvt. Ltd., India)

Click here to view


Currently, there are four different types of balloon-expandable valves available for the pulmonic position: the Melody®, the Sapien® S3, Inovare® (Braile Biomedica, Sao Jose do Rio Preto, Brazil), and the Myval®.

The next generations of valves are self-expandable Harmony TPV (Medtronic Inc, Minneapolis, Minn) and Venus P valve (Venus Medtech, Shanghai, China) in large RVOT's.[17],[18],[19]

According to the latest meta-analysis study published, while comparing with balloon-expanding valve (BEV), the self-expanding valve (SEV) was associated with better postprocedural effective orifice area, which reduces the risk of patient–prosthesis mismatch.[20] However, SEV was associated with higher rates of postprocedural pacemaker implantation and it stents the RVOT and thus prevents RV re-modelling.

The real challenge in BEV is to choose a proper valve size in native RVOT without an obstruction. We assessed the valve area by MSCT/angiographic balloon sizing and compared it with an area of the valve finally implanted [Table 3]. Between MSCT scan and balloon sizing, we felt that the balloon sizing gives the true narrowest diameter and increasing this area by 23%–25% will give the appropriate valve size for successful implantation. In fact, we increased the nominal volume by 5 ml to give an inner diameter of 33 mm for patient #3, which was based on the previous two valve experiences. The specific risks involved in oversizing are:
Table 3: Valve size assessment by comparing angiogram balloon sizing and multislice computed tomography measurements

Click here to view


Rupture of pulmonary artery

In a dilated native PA without circumferential calcification, this is unlikely as the elasticity of the native PA is good

Coronary obstruction

The balloon sizing is 23%–25% less in the area compared to the final valve area. Hence, a coronary occlusion on the final diameter cannot be ruled out. We recommend a selective coronary wire placement in the nearest coronary artery to the landing zone.

Limitations

This is a retrospective study. The procedural safety must be proven in further larger studies. Long-term follow-up data are required to assess the durability.


   Conclusion Top


TPVI is feasible in large native PA with or without obstruction using the new Myval®. Defining the landing zone by MSCT scan is crucial. We recommend a 23%–25% increase in the balloon sizing area to achieve a stable valve size.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Eicken A, Ewert P, Hager A, Peters B, Fratz S, Kuehne T, et al. Percutaneous pulmonary valve implantation: Two-centre experience with more than 100 patients. Eur Heart J 2011;32:1260-5.  Back to cited text no. 1
    
2.
Hager A, Schubert S, Ewert P, Søndergaard L, Witsenburg M, Guccione P, et al. Five-year results from a prospective multicentre study of percutaneous pulmonary valve implantation demonstrate sustained removal of significant pulmonary regurgitation, improved right ventricular outflow tract obstruction and improved quality of life. EuroIntervention 2017;12:1715-23.  Back to cited text no. 2
    
3.
Bonhoeffer P, Boudjemline Y, Saliba Z, Merckx J, Aggoun Y, Bonnet D, et al. Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction. Lancet 2000;356:1403-5.  Back to cited text no. 3
    
4.
Cabalka AK, Hellenbrand WE, Eicken A, Kreutzer J, Gray RG, Bergersen L, et al. Relationships among conduit type, pre-stenting, and outcomes in patients undergoing transcatheter pulmonary valve replacement in the prospective North American and European melody valve trials. JACC Cardiovasc Interv 2017;10:1746-59.  Back to cited text no. 4
    
5.
Auzina L, Lubaua I, Ligere E, Lace I, Bergmane I, Matsone-Matsate B, et al. Initial experience with Edwards SAPIEN valve transcatheter implantation in native RVOT in Latvia. Acta Med Litu 2020;27:10-6.  Back to cited text no. 5
    
6.
Plessis J, Hascoët S, Baruteau A, Godart F, Le Gloan L, Warin Fresse K, et al. Edwards SAPIEN transcatheter pulmonary valve implantation: Results from a French registry. JACC Cardiovasc Interv 2018;11:1909-16.  Back to cited text no. 6
    
7.
Chatterjee A, Bajaj NS, McMahon WS, Cribbs MG, White JS, Mukherjee A, et al. Transcatheter pulmonary valve implantation: A comprehensive systematic review and meta-analyses of observational studies. J Am Heart Assoc 2017;6:e006432.  Back to cited text no. 7
    
8.
Tanase D, Georgiev S, Eicken A, Ewert P. The Sapien valve provides enough grip to be implanted in pulmonary position without a pre-stent. Cardiovasc Diagn Ther 2019;9:S264-8.  Back to cited text no. 8
    
9.
Suntharos P, Prieto LR. Percutaneous pulmonary valve implantation in the native right ventricular outflow tract using a 29-mm Edwards SAPIEN 3 valve. World J Pediatr Congenit Heart Surg 2017;8:639-42.  Back to cited text no. 9
    
10.
Garay F, Pan X, Zhang YJ, Wang C, Springmuller D. Early experience with the Venus p-valve for percutaneous pulmonary valve implantation in native outflow tract. Neth Heart J 2017;25:76-81.  Back to cited text no. 10
    
11.
Sivaprakasam MC, Arvind A, Sridhar A, Gunasekaran S. First transcatheter pulmonary valve implantation – An Indian made valve. IHJ Cardiovasc Case Rep 2020;4:13-6.  Back to cited text no. 11
    
12.
Stout KK, Daniels CJ, Aboulhosn JA, Bozkurt B, Broberg CS, Colman JM, et al. 2018 AHA/ACC guideline for the management of adults with congenital heart disease: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. Circulation 2019;139:e698-800.  Back to cited text no. 12
    
13.
Available from: https://www.edwards.com/gb/devices/Heart-Valves/Transcatheter-Sapien-3 [Last accessed on 2021 Mar 20].  Back to cited text no. 13
    
14.
Bouzas B, Kilner PJ, Gatzoulis MA. Pulmonary regurgitation: Not a benign lesion. Eur Heart J 2005;26:433-9.  Back to cited text no. 14
    
15.
Kim H, Sung SC, Kim SH, Chang YH, Lee HD, Park JA, et al. Early and late outcomes of total repair of tetralogy of Fallot: Risk factors for late right ventricular dilatation. Interact Cardiovasc Thorac Surg 2013;17:956-62.  Back to cited text no. 15
    
16.
Available from: https://www.merillife.com/medical-devices/vascular-intervention/heart-valves/tavr/myval [Last accessed on 2021 Mar 20].  Back to cited text no. 16
    
17.
Benson LN, Gillespie MJ, Bergersen L, Cheatham SL, Hor KN, Horlick EM, et al. Three-year outcomes from the harmony native outflow tract early feasibility study. Circ Cardiovasc Interv 2020;13:e008320.  Back to cited text no. 17
    
18.
The Medtronic Harmony Transcatheter Pulmonary Valve Clinical Study. NML Identifier: NCT 02979587. Available from: https://clinicaltrials.gov/ct2/show/NCT02979587. [Last accessed on 2020 March 12].  Back to cited text no. 18
    
19.
Morgan G, Prachasilchai P, Promphan W, Rosenthal E, Sivakumar K, Kappanayil M, et al. Medium-term results of percutaneous pulmonary valve implantation using the Venus P-valve: International experience. EuroIntervention 2019;14:1363-70.  Back to cited text no. 19
    
20.
Lee HA, Chou AH, Wu VC, Chen DY, Lee HF, Lee KT, et al. Balloon-expandable versus self-expanding transcatheter aortic valve replacement for bioprosthetic dysfunction: A systematic review and meta-analysis. PLoS One 2020;15:e0233894.  Back to cited text no. 20
    

Top
Correspondence Address:
Dr. Muthukumaran C Sivaprakasam
Department of Pediatric Cardiology, Apollo Children's Hospital, Chennai, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/apc.apc_62_21